IR310E staff station
Brands
 
Shopping Cart
0 items
 

IR310E staff station

IR310E staff station
Click to enlarge
Price: $191.53
Availability: In Stock *
Model: IR310E
Manufacturer: Tektone
Average Rating: Not Rated

FEATURES
The IR310E Staff Station provides reliable two-way,
hands-free communication between the nurse master
station and utility room, examination rooms, kitchen, or
any location where staff members may need to originate
calls or where cord sets are not required. Calls are placed
by use of a momentary push button. Solid-state lights on
the panel indicate that the call has been placed and that the
intercom is in use.
As an additional feature, the staff station is field
programmable to function as a patient station, if desired.
• UL® 1069 Listed
• Supervised
• Microprocessor design
• Solid-state multiplex circuitry
• Field programmable for staff or patient station
operation
• Programmable call priority
• All stations interchangeable
• Simplified multiplex wiring using plug-in type
connectors
• Attractive flame retardant (UL® 94V-0) plastic panel
SPECIFICATIONS
Dimensions: Height: 4.5" (114 mm)
Width: 6.375" (162 mm)
Depth: 2.75" ( 70 mm)
Projects: 0.5" (12.7 mm)
Construction: Flame retardant (UL® 94V-0) plastic
Finish: Light texture, white
Speaker: 2.5" × 2.5" (64 mm × 64 mm)
Connections: Plug-in type
Housing: Steel City #H3BD with Steel City #3GC
plaster ring or exact equal. Horizontal
mount.
Wiring: Multiplex
REQUIRED COMPONENTS
Tek-CARE®NC300II Nurse Call System
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Staff station shall be TekTone® IR310E, or approved
equal. Model IR310E shall include solid-state LSI/
microprocessor technology, combination yellow callplaced
indicator and call button, and combination red inuse
indicator and reset button.
The station shall utilize a 2.5" × 2.5" speaker/microphone
for accurate voice reproduction. The station shall be field
programmable to emulate a patient station, if so desired.
All station components shall be mounted on an attractive,
flame retardant (UL® 94V-0) plastic panel. Connections
shall be plug-in type.
The IR310E shall be UL® 1069 Listed.